Detector Pilatus3 X CdTe 2M
Beam position on corner of detector (quarter detector)
Minimum energy threshold 37.5 keV
Detector distance 1.5 m
Wavelength / Energy 0.165 Å / 75.0 keV
Beam size ⌀ 0.05–0.075 mm 
Focusing on detector via transfocator
Q-range 0.16–9.1 Å-1
d-spacing 40–0.69 Å
Peak resolution 0.004 Å-1 (first peak integral breadth for LaB6 660b)
Sample thickness 1 mm
Sample motion vibration plus active translation

 

 