Skip to content
English
  • There are no suggestions because the search field is empty.

SAXS—measurement specs

Detector Pilatus3 X CdTe 2M
Flight tube Placed along beam-path after sample to minimize air scattering background
Beam position on center of detector (full detector)
Minimum energy threshold 37.5 keV
Detector distance 7.1 m
Wavelength / Energy 0.165 Å / 75.0 keV
Beam size ⌀ 0.05–0.075 mm 
Focusing on detector via transfocator
Q-range 0.01–0.78 Å-1
d-spacing 630–8 Å
Sample thickness 1 mm
Sample motion vibration plus active translation
Calibration standard Silver behenate