Can you detect trace phases in my sample?

Yes, our high-resolution setup is excellent for detecting minor phases in multiphase samples. Typically, crystalline phases present at levels of < 0.5–1wt% can be detected. Factors that affect detection limits include:

  1. Crystallinity: Well-crystallized minor phases are easier to detect than poorly crystalline ones
  2. Peak overlap: Minor phases with distinctive peaks separated from the main phase are easier to identify
  3. Absorption contrast: Large differences in absorption between phases can affect detection limits
  4. Sample homogeneity: Well-mixed samples provide more reliable quantification of minor phases
 

For optimal trace phase analysis, please provide information about suspected minor phases when submitting your samples.